Back to Universal Prosthetic Kit

Universal Prosthetic Kit Manual

COMPATIBILITY CHART

 

1. INDICATION FOR USE

The Universal Prosthetic Kit is used for dental requirements in the area of dental surgery.

Any other form of use is ill advised and can be harmful. 

This is an instrument with a precise system of torque control allowing the screwing of prosthetic parts at a definite torque. 

 

2. AREA OF APPLICATION 

The Universal Prosthetic Kit must only be used by someone with dental experience, with knowledge in implantology as well as in prosthetic techniques related to dental implants for the application described here above, and in compliance with the provisions in effect in relation to the prevention of accidents at work, protection of employment, and the instructions for use.

 In accordance with these provisions, the user must only use working instruments without defects, observe the principle of correct use, protect themselves and the patient or third party against any danger, and prevent all contamination by the product. 

 

3. GENERAL SAFETY INFORMATION  

Before use, check that the Universal Prosthetic Kit is not damaged or that no part of it is missing. Use protective gloves, glasses, and a mask. 

It is recommended to the practitioner to test the torque wrench before using it on the patient in order to check that it is correctly functioning. In the event of visible malfunction or damage, immediately stop using the instrument and inform contact@conmetior.com.

 

4. TECHNICAL CHARACTERISTICS 

The user is responsible for the appropriate choice of operating conditions, which will prevent any risk to the patient or to another person. All the torque values are indicated on the bottom of the driver. In choosing the torque, follow the recommendations of the implant manufacturer.

 

5. HYGIENE AND MAINTENANCE

The re-sterilization of reusable medical devices must be performed by a properly trained and protected member of staff. 

The re-sterilization protocol must consider the risk of infection. Refer to the manufacturer’s instructions for each product used. 

Particular care to adhere to the instructions regarding concentrations, lengths of exposure, the replacement of solutions, and the lifespan of the products. 

Never combine products and adhere to the guidelines regarding the disposal of used products.  

Wear suitable protective clothing.

Never immerse the torque wrench in any kind of solution.  

Never clean the torque wrench in an ultrasonic tank.

Clean the device immediately after an operation.

Use only detergent or disinfectant solutions with a neutral or slightly alkaline PH.

It is extremely unadvisable to use products which may fix X proteins to the unit (e.g. alcohol, aldehydes...).

Do not use sodium hypochlorite (bleach) as there is a high risk of corrosion.

6.1 EXTERNAL CLEANING

The torque wrench and the case can be cleaned by using a brush under tap water. Do not autoclave the case.

External disinfection using a spray, or a wipe soaked with disinfectant products microbiologically controlled (ph. between 2.5-9). In any case, it is important to follow the reaction time of the disinfectant product used.  

 

STERILIZATION

Do not sterilize the case (refer to external cleaning instructions)

Read the instruction leaflet provided by the sterilizer manufacturer. 

Adhere to the space specified between pouches and do not overload the sterilizer.

Only sterilize instruments which have previously been disinfected, cleaned, lubricated, and tested, the case can not be sterilized.

Remove the rotating instrument from the torque wrench prior to sterilization.

Make sure that the device does not have any areas of corrosion or cracks, and check that it is operating properly.

Ensure that the product is dry; if necessary, dry any residual water with medical quality pressurized air. 

Use sterilization pouches suitable for the torque wrench and the sterilizer. Always use one single torque wrench per pouch.

In order to prevent any retention of water, place the pouch in the sterilizer in such a way that any concave parts are face down.

If the sterilizer has several types of cycles, choose a cycle designed for medical devices (a minimum of 130°C at 2.13 bars (266°F at 30.88 psi) for 18 minutes.)

After each sterilization cycle, check that there is no water remaining on the inside and outside of the packaging. Make sure that the flow indicator has changed to the correct color. 

Keep the devices in the sterilization pouches away from light, moisture, and any kind of contamination. 

Follow the manufacturer’s recommendations as seen on the packaging. The duration for which the device is kept after sterilization should not exceed 1 month. Label the devices, specifying the expiration date. After the expiration date, repeat the cleaning and sterilization procedure. 

 

GUARANTEES

If you think you have received a product in poor condition or if you think the packaging has been tampered with or damaged before delivery, please refuse to accept the package and return it to the delivery person.

Email us at contact@conmetior.com mentioning your order ID. We will personally ensure that a brand-new replacement is issued to you with no additional cost.

Make sure that the ORIGINAL PRODUCT TAG is present when you send back the product.

You must return any goods within 10 days of date of receipt.

Due to Federal regulations, apart from condition reserved herein above all sales for any of our products are FINAL and NON-REFUNDABLE.

This Universal Prosthetic Kit is guaranteed against all manufacturing defects for 6 months from the date of invoice. This guarantee does not apply to wear and tear parts. 

All changes or additions to the product without the express agreement of Conmetior render this guarantee null and void. 

The guarantee becomes null and void if the technical instructions are not followed. 

Conmetior cannot be held responsible for damage resulting from or which could result from normal wear, use, cleaning or incorrect maintenance, the non-observance of instructions for use or connection, scaling or corrosion, impurities in the water supply system or unusual chemical or electrical influences or non-observance of the instructions, maintenance instructions and assembly of Conmetior and other manufacturer’s instructions. 

Delivery charges incurred when sending an instrument back to Conmetior for repair will be paid by the client, even if the repair itself is covered by the guarantee. 

Postage and packaging fees when returning the instrument to the client are covered by the guarantee. 

So that guarantee requests are taken into consideration, please attach a copy of the invoice or a copy of the delivery slip.

Compatibility Chart

The chart is organized by Driver Type.

IMPLANT COMPANIESCOUNTRYDRIVER TYPECompatible Universal Prosthetic KitCompatible with Right Angle Driver
GMI PhoenixSPAIN0.031"-0.80mm No
Bredent Implant GmbHGERMANY0.035"-0.9mmYes
Medentis (t-ecco femaleinner part and ICX-mini cover screw)GERMANY0.035"-0.9mmNo
Neodent CM final abutmentBRAZIL0.035"-0.9mmYes
VictoryFRANCE0.035"-0.9mmYes
Sterngold Implamed Dental Implant SystemsUSA0.039"- 1.0 mm Yes
Dentsply AnkylosGERMANY0.039"-1.0mm yes
T.B.R GroupFRANCE0.042"-1.09 mm Yes
IDI 4051FRANCE0.046"-1.17mmno
ANTHOGYRFRANCE0.047"-1.19mmYes
Odontit S.AARGENTINA0.047"-1.2mmNo
ArgonGERMANY0.047"-1.2mmYes
Biotech dentalFRANCE0.047"-1.2mmYes
Drive implantsFRANCE0.047"-1.2mmYes
General ImplantsGERMANY0.047"-1.2mmYes
Gunter medicalGERMANY0.047"-1.2mmYes
Medentis (ICX-Hexagon instrument)GERMANY0.047"-1.2mmYes
Medical instinctGERMANY0.047"-1.2mmYes
Neodent CM everything elseBRAZIL0.047"-1.2mmYes
VictoryFRANCE0.047"-1.2mmYes
BiolineISRAEL0.048"-1.22mmYes
BlueSkyBio MaxKOREA/USA0.048"-1.22mmYes
Cowell MediSOUTH KOREA0.048"-1.22mmNo
GMI AvantguardSPAIN0.048"-1.22mmYes
GMI FrontierSPAIN0.048"-1.22mmYes
GMI MonolithSPAIN0.048"-1.22mmNo
GMI PhoenixSPAIN0.048"-1.22mmYes
Hiossen/Osstem Comp. LtdSOUTH KOREA0.048"-1.22mmYes
Keystone/Lifecore RenovaUSA0.048"-1.22mmYes
Keystone/Lifecore RestoreUSA0.048"-1.22mmYes
MegaGenSOUTH KOREA0.048"-1.22mmYes
Neobiotech Comp. LtdSOUTH KOREA0.048"-1.22mmYes
Southern Implant ExternalSOUTH AFRICA0.048"-1.22mmYes
BTI Biotechnology Institute, S.L.SPAIN0.048"-1.22mm Yes
Dentsply Friadent Ceramed Inc, USA / Friadent GmbH GERMANY0.048"-1.22mm Yes
Sterngold Implamed Dental Implant SystemsUSA0.048"-1.22mm Yes
ConmetRUSSIA0.049"-1.25mmYes
ACE Internal Hex implantsUSA0.050"-1.25mmYes
AIDI biomedical USA0.050"-1.25mmYes
ALFA-GATEISRAEL0.050"-1.25mmYes
AMERICAN DENTAL IMPLANT CORPORATIONUSA0.050"-1.25mmYes
ARDS IMPLANTISRAEL0.050"-1.25mmYes
ASTRA TECHSWEDEN0.050"-1.25mmYes
Aurosan dentalGERMANY0.050"-1.25mmYes
BEGO GmbHGERMANY0.050"-1.25mmYes
BioHorizonsUSA0.050"-1.25mmYes
BlueSkyBio Conus 12 & ThreeKOREA/USA0.050"-1.25mmYes
BlueSkyBio internal Hex & MolarKOREA/USA0.050"-1.25mmYes
Btlock s.r.l.ITALY0.050"-1.25mmYes
Camlog GroupSWISS0.050"-1.25mmNo
CortexISRAEL0.050"-1.25mmYes
Dental TechITALY0.050"-1.25mmYes
DentiumSOUTH KOREA0.050"-1.25mmYes
Ditron DentalISRAEL0.050"-1.25mmYes
Dr Ihde Dental, GmbHGERMANY0.050"-1.25mmyes
Eckermann LaboratoriumSPAIN0.050"-1.25mmYes
Hi-Tec ImplantsISRAEL0.050"-1.25mmYes
IDI 4041FRANCE0.050"-1.25mmYes
Impladent LtdUSA0.050"-1.25mmYes
Implant LogisticsSPAIN0.050"-1.25mmNo
Implant VELARGENTINA0.050"-1.25mmYes
ImplantdirectUSA0.050"-1.25mmYes
IMTEC CorporationUSA0.050"-1.25mmYes
Innova LifeSciences Co, (Sybron Dental from 2006)CANADA0.050"-1.25mmNo
Intra-Lock System InternationalUSA0.050"-1.25mmNo
IQ ImplantISRAEL0.050"-1.25mmYes
LASAK LtdCZECHIA0.050"-1.25mmNo
MIS (Medical Implant System)ISRAEL0.050"-1.25mmYes
Nanoplant HochstGERMANY0.050"-1.25mmYes
Neobiotech Comp. LtdSOUTH KOREA0.050"-1.25mmYes
OCO BIOMEDICALUSA0.050"-1.25mmYes
Oral implant S.r.lITALY0.050"-1.25mmYes
PaltopISRAEL0.050"-1.25mmYes
PaltopUSA0.050"-1.25mmNo
RadixGERMANY0.050"-1.25mmYes
Sterngold Implamed Dental Implant SystemsUSA0.050"-1.25mmYes
Sweden&Martina SpAITALY0.050"-1.25mmYes
TFI SystemITALY0.050"-1.25mmYes
TRI dental implantsSWISS0.050"-1.25mmYes
Zimmer DentalUSA0.050"-1.25mmNo
ADIN-IMPLANTSISRAEL0.050"-1.27mm Yes
Bone System implantITALY0.050"-1.27mm Yes
Ritter implantsGERMANY0.051"-1.29mmYes
CumdentGERMANY0.051"-1.3mmYes
Medentis (for all abutments screws)GERMANY0.055"-1.4mmNo
NDI medicalGERMANY0.055"-1.4mmYes
Neodent GMBRAZIL0.055"-1.4mm Yes
AB DENTAL IMPLANTISRAEL0.055"-1.4mm Yes
Bone System implantITALY0.059"-1.5mmYes
ALPHA BIO IMPLANT INCISRAEL0.060"-1.5mmYes
Implant Microdent SystemSPAIN0.063"-1.60mmYes
SybronUSA0.067"-1.7mm Yes
Sterngold Implamed Dental Implant SystemsUSA0.080"-2.0mm Yes
Bio implants Standard plateformGERMANY0.083"-2.1mmYes
TOV implantsISRAEL0.083"-2.1mmNo
ALPHADENT BIO HPIDKGERMANY0.090"-2.35mmYes
JDENTALCAREITALY0.095"-2.42mmNo
Bio implants wide plateformGERMANY0.098"-2.5mmYes
Dyna Dental EngineeringHOLLAND0.098"-2.5mmYes
ALPHADENT IDKGERMANY0.120"-3.0mmNo
Conmet (prior to 2020)RUSSIA1.0mm squareNo
Thommen MedicalSWISSFour LobeYes
Keystone/Lifecore Genesis & PrimaUSASquareYes
BlueSkyBio One StageKOREA/USAStar &0.048"-1.22mmYes
Klockner ImplantsSPAINStar torkYes
BlueSkyBio QuatroKOREA/USAStar/TorxYes
Southern Implant OctaSOUTH AFRICAStar/TorxYes
StraumannSWISSStar/TorxYes
ACE Octagon implantsUSAStar-gripYes
ACE Tri-lob implantsUSAUnigripYes
BranemarkSWEDENUnigripNo
Glidedentalimplant HahnUSAUnigripYes
NeossU.KUnigripNo
Nobel/BranemarkSWEDENUnigripNo
Noris MedicalISRAELUnigripYes
Southern Implant Tri NexSOUTH AFRICAUnigripYes
BlueSkyBio trilobeKOREA/USAUnigrip & 0.048" Yes
OEM Dental ImplantsHOLLANDunigrip. not for NobelNo